R6243/6244 DC Voltage Current Source/Monitor

DC Voltage Current Source/Monitor (NOT distributed in North America & Japan.)

Features
  • Voltage source and current measurement (VSIM)/
  • Current source and voltage measurement (ISVM)
  • Voltage source and voltage measurement (VSVM)/
  • Current source and current measurement (ISIM)
  • Resolution up to 1µV and up to 100pA
  • Sink enabled bipolar output
  • Minimum pulse width: 1 ms
  • Linear, logarithmic, and random sweep functions for characteristics tests
  • Limiter (compliance), oscillation, overload, and overheat detection functions
  • Synchronous operation function by combining two R6243/6244s or more
  • GPIB for automatic measuring system

R6243 R6244
Maximum output curren ± 2.0 A at up to ± 32 V
± 1.0 A at up to ± 64 V
± 0.5 A at up to ± 110 V
±10 A at up to ±7 V
±4 A at up to ± 20V
Voltage source/measurement range 320 mV to 110 V 320 mV to 20 V
Current source/measurement range 32 µA to 2 A 320 µA to 10 A
Number of digits
    Source
    Measurement
4 1/2
5 1/2
Voltage source/measurement resolution
Source
Measurement
10 µV
1 µV

Measurement
1 nA
100 pA
10 nA
1 nA